추천 제품
기술
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
검사 파라미터
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: extract from 50 g homogenized fruit or vegetable sample
SPE tube/cartridge: ENVI-Carb, 500 mg/6 mL (57094)
sample addition: 0.5 mL concentrated sample extract
elution: 20 mL acetonitrile:toluene (3:1)
eluate post-treatment: Concentrate to 2 mL by rotary evaporation. Add 2 x 10 mL acetone concentrating to 2 mL after each acetone addition. (Add 50 μL I.S. (500 ng/μL cis-chlordane in acetone) then dilute to 2.5 mL with acetone.)
column: 14% cyanopropylphenyl polysiloxane, 30m × 0.25 mm I.D., 0.15 μm (Supelco equivalent, Equity-1701, 28372-U)
oven: 70 °C (hold 2 min) to 130 °C at 25 °C; 130 °C to 220 °C at 2 °C/min; 220 °C to 280 °C at 10 °C/min and hold 4.6 min
detector: MSD, 285 °C
carrier gas: helium
injection: 2 μL, splitless
적합성
application for SPE
응용 분야
agriculture
food and beverages
애플리케이션
Homogenize 50 g chopped fruit/vegetable sample in 100 mL acetonitrile. Add 10 g sodium chloride and homogenize 5 min. Transfer 13 mL of acetonitrile layer to clean centrifuge tube. Centrifuge at high speed for 5 min. Evaporate 10 mL aliquot of supernatant to 0.5 mL under Nitrogen at 35°C.
법적 정보
가장 최신 버전 중 하나를 선택하세요:
Chromatograms
application for HPLC자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..