GC Analysis of Volatiles in Purple Nitrile Gloves on SPB®-HAP after Collection/Desorption using Carbotrap® 300 Tube

GC Analysis of Volatiles in Purple Nitrile Gloves on SPB®-HAP after Collection/Desorption using Carbotrap® 300 Tube suitable for GC, application for air monitoring

Materials

관련 제품

제품 번호
설명
가격

ATIS Adsorbent Tube Injector System 110V

ATIS Adsorbent Tube Injector System 230V

ATIS Extraction Glassware with Micro Connector

CONDITIONS

sample/matrix

0.5 g purple nitrile glove

thermal desorption tube

Carbotrap® 300 Glass-Fritted Thermal Desorption Tube, 1/4 in O.D. x 3.5 in L (30141-U)

desorption process

350 °C with 10 min hold

transfer line/valve temp.

180 °C

column

SPB®-HAP, 60 m x 0.32 mm I.D., 4 μm (25020-U)

oven

40 °C (3 min), 12 °C/min to 200 °C (20 min)

detector

MSD

MSD interface

210 °C

carrier gas

helium

dry purge

15 min

설명

일반 설명

Qualitative volatile headspace outgassing from purple nitrile gloves using the ATIS system with analysis by thermal desorption.