GC Analysis of an 8-Component Solvent Mix in Water on Watercol™ 1910 suitable for GC

Materials

Analytical column

Product No.
Description
Pricing

Watercol 1910 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

Watercol 1910, 30 m x 0.25 mm I.D., 0.20 μm (29711-U)

oven

35 °C, 4 °C/min to 125 °C (2 min)

inj. temp.

250 °C

detector

TCD, 300 °C

carrier gas

helium, 25 cm/sec

injection

1 μL, 100:1 split

liner

4 mm I.D., split type, cup design

sample

8-component solvent mix in water

Description

Analysis Note

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water does not interfere chromatographically with other small polar analytes. This application shows the analysis of an 8-component solvent mix in water on Watercol 1910.

Legal Information